Abstract
IR transmitting AgCl xBr1- x, fibers were used as internal reflection elements for the spectroscopic detection of hydrocarbons that are dissolved in water in an attenuated total reflection (ATR) type of setup. The fibers were partly flattened by pressing and then coated with an analyte-enriching organic polymer in order to increase the sensor sensitivity. A portable tunable diode laser spectrometer served as a spectrometric unit. The dependence of the sensitivity on coupling conditions was investigated and lead to an optimized coupling. It is shown that the sensitivity of the flattened waveguide with a thickness of 170 μm can be enhanced by a factor of 5 compared to a fiber with 900 μm diameter. Aqueous toluene solutions with concentrations in the parts-per-million range were used to characterize the sensor system that showed a linear response with regard to analyte concentration. The noise equivalent concentration (NEC) was on the order of 1 ppm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.