Abstract

Recently, an assay for detection of provirai HIV-1 DNA in leukocytes became commercially available. This assay (Amplicor HIV-1 test, Roche Diagnostic Systems) multiplies HIV-1 DNA up to a detectable level, using the polymerase chain reaction. We studied performance of this assay on 74 samples from HIV-1-infected patients and on 41 samples from healthy blood donors. Twice a negative control sample appeared to be erroneously reactive. However, sensitivity and specificity on the patient and donor samples both were 100%. To avoid false-positive results, we advise to repeat initially reactive samples if no other data confirm HIV-infection.