Abstract

A very simple and inexpensive method to detect oversulfated chondroitin sulfate (OSCS) or other high-charge density polyanionic structures as contaminants in heparin products using potentiometric polyanion sensors is described. In the potentiometric measurement, a greater change in the phase boundary equilibrium potential of polymeric membranes formulated with tridodecylmethylammonium (TDMA) anion exchange sites is observed for the contaminated heparin than for the untainted heparin due to the higher charge density of OSCS or other impurities compared to that of porcine heparin. Detection of 0.5 wt % OSCS impurity is readily achieved using only 1 mg/mL of final polyion concentration. Even lower detection limits for OSCS contamination may be possible if higher final concentrations of heparin preparations can be employed in the test procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.