Abstract

This study is the first reported ion mobility detection of ethyl centralite and diphenylamine (DPA) smokeless gunpowder stabilizers, together with the nitroso and nitro derivatives of diphenylamine. First, the applicability of the ion mobility spectrometry (IMS) for the substances of interest was determined. The existence of numerous peaks, both in positive and negative modes, clearly demonstrates the success of these experiments. All mono and di-nitro derivatives of DPA tested were detected with this method. Unfortunately, many of the ions generated were not accurately identified. However, reduced mobility constants representative of each ion generated under defined operating conditions could be used for purpose of compound identification.The method was then successfully tested on real gunpowder samples.By the use of IMS, we managed to establish a rapid, simple and sensitive screening method for the detection and identification of smokeless gunpowder organic components.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.