Abstract

Fast and reliable detection of endotoxin (ET) in medical equipment and pharmaceutical products is an essential precursor to clinical treatment. In this study, we demonstrate the use of shifts in wavelength of photonic crystal nanolasers for sensing the Limulus amebocyte lysate reaction, which is a standard method for detecting ET. From working curves of wavelength shift vs ET concentration, whose correlation factors were as high as 98%, we detected a required concentration of 0.001 EU/ml within 33 min and detected a low concentration of 0.0001 EU/ml.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.