Abstract

The establishment of a convenient and effective detection method for doxycycline (DC) holds significant importance in drug monitoring and drug residue assessment. In this work, carbon quantum dots (CQDs) with excellent and stable luminescence performance (the quantum yield of CQDs was 21.8%) were synthesized by the melting method and employed as probes to monitor the fluorescence intensity variations before and after the introduction of DC. A fluorescence analytical method based on the internal filtration effect (IFE) was developed for DC determination. The mechanism of DC quenching CQDs was verified using fluorescence lifetime tests, absorption spectroscopy, and evaluation of internal filtration parameters. After optimizing experimental conditions, it was found that the DC concentration (CDC) exhibited a good linear relationship with the fluorescence quenching efficiency ((F0-F)/F0) of CQDs in the range of 5-30 µM. The fitted linear equation was Y = 0.01249*CDC+0.03625, R2 = 0.9987, and the detection limit was 2.343 µM (n = 8). This developed method has been successfully applied to accurately determine DC concentrations in both doxycycline hydrochloride tablets and human serum samples. It stands out for its simplicity, rapidity, and acceptable detection performance. Due to its advantages, this method holds great promise for application in the biomedical field for monitoring DC drug concentrations and ensuring quality control.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call