Abstract
The non-structural 1 (NS1) protein of the dengue virus circulates in infected patients’ blood samples and can be used for early diagnosis of dengue infection. In this paper, we present the detection of naturally-occurring dengue NS1 antigen in infected patient blood plasma using straight long-range surface plasmon waveguides. Three commercially-available anti-NS1 monoclonal antibodies were used for recognition and their performance was compared and discussed. A similar figure of merit to the one used in conventional dengue NS1 capture using an enzyme-linked immunosorbent assay (ELISA) was applied to our results. In general, the positive patient samples can be clearly differentiated from the negative ones and the results agree with those obtained using ELISA. The largest signal-to-noise ratio observed during the experiments was 356 and the best detection limit observed is estimated as 5.73pg/mm2.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.