Abstract

A differential optical absorption spectroscopy (DOAS) gas-analyzer was successfully tested. A high pressure 150-W Xe arc lamp was employed as a light source. This system consisted of a coaxial telescope, a spectrometer, an analyzer and retroreflector. In order to record the spectra, a monochrometer with a grating and photodiode array was adopted. Gas analyzer spectral data bank includes more than 30 moleculas absorbed in UV spectral region. The measured absorption spectra were evaluated by using a least-squares to determine the average mixing ratio of each species in the atmosphere. A number of air pollutants concentrations: SO2, NO2, O3, etc were trace measured. Minimally detected concentration on pathlength 400 m is the unit of ppb at the time of accumulation of 2 min. The results of the field test measurements of pollutants in Tomsk are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.