Abstract
A method for using charge injection devices (CIDs) for detection of high-energy charged particles from inertial-confinement fusion reactions is described. Because of the relatively small depletion region of the CID camera (depletion depth of approximately 7 mum), aluminum foils are placed in front of the device to reduce the energy of the charged particles and maximize the energy deposited in the CID. Simultaneous measurements of (2)H(d,p)(3)H protons with a CID and a surface barrier detector indicate that the CID is an efficient detector of charged fusion products. Tests using high energy alpha particles emitted from a radium-226 source are also reported.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.