Abstract

Probe electrospray ionization (PESI) is a recently developed ionization technique which uses a solid needle or wire as sampling probe and ESI emitter instead of capillary. PESI is free from the clogging problem and it has high tolerance to salts and urea. We present herein a comparative study of the probe electrospray ionization (PESI) and nano-electrospray ionization (nano-ESI) for the measurement of biomolecules in the sample solutions with high concentrations of salts and urea. Our results show that PESI could provide equivalent ionization performance with nano-ESI, and in certain cases it could be superior to nano-ESI for the samples with high concentration (>100 mM) of salts and urea. Therefore, PESI can be useful for the direct analysis of cell/tissue extracts and protein digestion without or reduced procedures in sample purification.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.