Abstract

To compare the accuracy of approximal caries detection using enhanced and unenhanced storage phosphor images and dental X-ray film. Fifty premolar/molar teeth were exposed simultaneously under standardized conditions on storage phosphor (SP) image plates (Digora) and Ektaspeed film. The image files were transferred to a different platform for image processing. Using a five-point confidence scale, ten observers rated 95 approximal surfaces for caries in enamel and dentin with three image modalities: enhanced and unenhanced SP images and film. The observations were validated microscopically. Diagnostic accuracy was measured as the area beneath the ROC curve (Az). Enhanced SP images had significantly higher Az values in both enamel (Az = 0.819) and dentin (Az = 0.845) than unenhanced images and film. No significant differences were found in either enamel or dentin between unenhanced SP images (Az = 0.780 and Az = 0.789) and film (Az = 0.688 and Az = 0.793). Enhancement of storage phosphor images improved detection of approximal caries in this in vitro study compared with unenhanced images and Ektaspeed film.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.