We demonstrated the detection of airborne bacteria by a disposable bio-precipitator and NanoGene assay combination. The bio-precipitator employed micro corona discharge at 1960V and at less than 35µA to simultaneously charge, capture and lyse the airborne bacteria. This was enabled by the use of a 15μL liquid anode. Using a custom exposure setup, the target bacterium Bacillus subtilis in the atomization solution was rendered airborne. After exposure, the liquid anode in the bio-precipitator was subsequently measured for DNA concentration and analyzed with the NanoGene assay. As the bacterial concentration increased from 0.0104 to 42.6 g-DCW/L the released DNA concentration in the liquid anode increased from 2.10±1.57 to 75.00±7.15ng/μL. More importantly, the NanoGene assay showed an increase in normalized fluorescence (gene quantification) from 18.03±1.18 to 49.71±1.82 as the bacterial concentrations increased from 0.0104 to 42.6 g-DCW/L. the electrical power consumption of the bio-precipitator was shown to be amenable for portable use. In addition, the detection limit of bio-precipitator and NanoGene assay combination in the context of environmentally relevant levels of airborne bacteria was also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.