Abstract

Detection of acetylcholine receptor (AChR) blocking antibodies through the use of a radiolabel has become standard procedure in most laboratories. Known drawbacks associated with radioassay, including cost of radioisotopes, hazards to laboratory professionals, and manufacture and disposal of radioactive materials, have prompted investigation into replacement assays. We describe here a high-throughput immunofluorescent flow cytometric assay designed for the detection of AChR blocking antibodies. In total, 323 serum samples were tested on both the AChR blocking radioassay and the new immunofluorescent flow cytometric assay. Analysis of the results revealed a 96.9% concordance between the two assay methods. Our results indicate that a new immunofluorescent flow cytometric AChR blocking antibody assay is not only feasible but clinically comparable in both sensitivity (91%) and specificity (99%) compared with radioassay.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.