Abstract

Mastitic milk was detected by obtaining conductance measurements using an impedance microbiology Bactomatic 120 SC instrument. Conductance readings separated normal and abnormal milks after 30 min at 25°C when readings differed by more than 2 to 3% and exceeded the variance among instrument module wells. Samples blended from four quarters of a cow indicated milk from one quarter was abnormal if the salt level in the abnormal quarter raised the blend conductivity above that of normal samples and variance among the wells. Either solid or liquid substrates that contained stimulants could be used to accelerate bacterial acid production or reduce impedance detection times and did not affect the ability to detect abnormal milk. However, measurements varied with the volume of sample in the well, suggesting that fixed 1-ml liquid volumes of milk be used. Such volumes would allow detection of abnormal milk and bacterial load on the same sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.