Abstract
A low noise UV-enhanced silicon photodiode has been modified into a low energy particle detector by removing the antireflection oxide layer. The electronic noise of a 100 mm2 device and a conventional charge sensitive preamplifier has been measured to 2.3 keV FWHM at room temperature operation. The response of this detector to low energy particles has been investigated. Electrons with 10 keV initial energy looses 0.8 keV in the surface deadlayer. Further, the sum of the energy losses in the deadlayer and to nuclear collisions have been measured to about 3.8 keV for 10 keV protons and 5.4 keV protons and 5.4 keV for 12 keV He+ ions
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.