Abstract
The method of detecting trace 1,4-dioxane in water using membrane extraction coupled with gas chromatography/differential mobility spectrometry (GC/DMS) was developed. The parameters including radio frequency voltage, sampling flow rate, permeation time and trapping time were optimized to 1000 V, 50 mL/min, 30 min and 150 s, respectively. The linear range for dioxane was obtained from 2.0 μg/L to 20.0 μg/L. The LOD was found to be 0. 67 μg/L. The specificity towards 1,4-dioxane in the presence of five chlorinated hydrocarbons was improved by using two-dimensional GC separation with optimized DMS compensation voltage. This method paves a way for developing field-deployable sensors for real-time monitoring contaminants in groundwater.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.