Abstract

Access openings and lateral root perforations were prepared in 40 single-rooted extracted teeth. The distance to the perforations was measured visually and then with the Tri Auto ZX. The electronic device was used first in its apex-locating function and then in conjunction with its automatic reverse mechanism. On average, the electronically measured length was slightly shorter than the visually measured length (P < 0.05). When the automatic reverse mechanism was used, the values registered were either consistent with or slightly different (+/- 0.5 mm) from the electronically measured length (P > 0.05). The results of this study revealed that the Tri Auto ZX detected and measured endodontic root perforations within a range of clinically acceptable variations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.