Abstract

Performance characteristics of two commercial quantitative Hepatitis A virus (HAV) RT-PCR assays, the LightCycler Hepatitis A virus quantification kit (Roche Diagnostics) and the RealArt HAV LC RT PCR kit (artus GmbH) for detection and quantification of HAV were evaluated. Both assays rely on reverse transcription and amplification of extracted RNA. Dilutions of two HAV strains, belonging to different subtypes, were prepared to determine the precision, accuracy, linearity and the detection limit. Both assays were found to be suitable for quantification measurement of HAV RNA, but only the Roche kit was able to distinguish the different HAV strains tested. The linear range for the artus assay was 10 4–10 6 IU/ml and 2 × 10 4 to 2 × 10 8 RNA copies/ml for the Roche assay. The detection limit of Roche kit was 2 TCID 50/ml or 500 RNA copies/ml and 5 TCID 50/ml or 50 IU/ml for the artus kit. Despite these small differences it is concluded that both assays are very suitable for detection and quantification of most prevalent HAV subtypes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.