Abstract
An error separation method called three-rosette method allows all errors in rotary tables integrated into coordinate measuring machines (CMMs) to be determined by means of an uncalibrated circular ball plate. The calibration procedure takes advantage of the fact that error contributions can be separated and eliminated by means of multiple measurements of each ball center in well-defined measurement positions. The skilled evaluation of these results allows all six degrees of freedom (DoF) of a rotary table to be detected: the angular positioning error motion, the axial error motion and two error components for describing both the tilt error motion and the radial error motion. The method is of practical and economic relevance, as ball plates can be manufactured at low cost and require no further calibration. A detailed measurement and evaluation strategy as well as example results are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.