Abstract

This paper presents our detailed design studies on capacitive micromachined ultrasound transducers (CMUTs) via finite element models developed in ANSYS 12. In this study, we will discuss our experimentally verified coupled field finite element simulations for operation frequency and bandwidth. The effects of device geometry, type of coupling media, material selection and properties are investigated for the frequency response, and bandwidth using an analytical approach. A commercial design of experiment software, Minitab is utilized for sample runs for obtaining response curves. The design principles for CMUTs are given in detail for effective and quick design. The design guidelines illustrate that, the frequency response is a strong function of transducer geometry, material selection and coupling media.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call