Abstract
The collection of a low vapor pressure chemical simulant triethyl phosphate sorbed onto silica gel (TEP/SG) from a surface with subsequent analysis of the TEP/SG particulates using desorption electrospray ionization–mass spectrometry (DESI–MS) is described. Collection of TEP/SG particulates on a surface was accomplished using a sticky screen sampler composed of a stainless steel screen coated with partially polymerized polydimethylsiloxane (PDMS). DESI–MS analysis of TEP/SG particulates containing different percentages of TEP sorbed onto silica gel enabled the generation of response curves for the TEP ions m/z 155 and m/z 127. Using the response curves the calculation of the mass of TEP in a 25wt% sample of TEP/SG was calculated, results show that the calculated mass of TEP was 14% different from the actual mass of TEP in the sample using the m/z 127 TEP ion response curve. Detection limits for the TEP vapor and TEP/SG particulates were calculated to be 4μg and 6 particles, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.