Abstract

In this article, a comprehensive uncertainty analysis will be given of a popular 2D particle measuring technique called Interferometric Laser Imaging for Droplet Sizing (ILIDS) for measuring droplets and bubbles. This includes the first estimates on the influence of droplet non-sphericity. Next to this, the uncertainty on the frequency determination and the calibration procedures were also investigated. To this end, the two most common calibration procedures were experimentally compared. This showed that it often dominates the total uncertainty. Design guidelines are then derived concerning observation angle, out-focus disk size and calibration procedure which could lead to a combined uncertainty below 1.5% for droplets and 2% for bubbles.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.