Abstract

Simulations show that significant improvement in imaging performance can be achieved through collimator design for thermal and fast neutron radiography with a laboratory neutron generator. The radiography facility used in the measurements and simulations employs a fully high-voltage-shielded, axial D–D neutron generator with a radio frequency driven ion source. The maximum yield of such generators is about 1010 fast neutrons per seconds (E = 2.45 MeV). Both fast and thermal neutron images were acquired with the generator and a Charge Coupled Devices camera. To shorten the imaging time and decrease the noise from gamma radiation, various collimator designs were proposed and simulated using Monte Carlo N-Particle Transport Code (MCNPX 2.7.0). Design considerations included the choice of material, thickness, position and aperture for the collimator. The simulation results and optimal configurations are presented.