Abstract
A low background thermal neutron flux detection system has been designed to measure the ambient thermal neutron flux of the second phase of the China Jinping Underground Laboratory (CJPL-II), right after completion of the rock bolting work. A 3He proportional counter tube combined with an identical 4He proportional counter tube was employed as the thermal neutron detector, which has been optimised in energy resolution, wall effect and radioactivity of construction materials for low background performance. The readout electronics were specially designed for long-term stable operation and easy maintenance in an underground laboratory under construction. The system was installed in Lab Hall No. 3 of CJPL-II and accumulated data for about 80 days. The ambient thermal neutron flux was determined under the assumption that the neutron field is fully thermalized, uniform and isotropic at the measurement position.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.