Abstract

ABSTRACT Design of the fluoride ion-selective electrode (FISE) as the tubular detector used in the flow injection analysis (FIA) has been described. Among other things, this design makes it possible to use various internal contacts. The effect of pH on peak height and detection limit in the pH range from 2.8 to 8.0 has been examined. The effect of flow rate and sample injection volume on peak height and range of linear response has also been examined. The flow rates range from 0.56 mL/min to 4.05 mL/min, while the injection volumes are 100 and 200 μL. The optimum conditions are carrier solution (0.1 M KNO3 buffer pH 4 and 10-6 M NaF), flow rate 1.54 mL/min and sample injection volume 100 μL. Applicability of the FISE as the tubular detector in determination of Fe(III)-ions and AI(III)-ions by flow infection analysis has been examined at pH values of 2.8 and 3.4.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.