Abstract
The thermocouple calibration temperature control system is constructed USB - 4716 200 ks/s, 16 multi-function USB module, JKH-C series thyristor phase shift trigger, PC machine as the main hardware, the U.S. National Instrument (NI) company Labview as the software development platform. According to the precious metal, base metal thermocouple wire thermoelectric EFM method to determine the temperature test point, the temperature control system using automatic control to test furnace temperature, and realize the visual operation interface. Temperature control accuracy: temperature deviation calibration point is less than ± 5 °C when testing the thermocouple, constant temperature stability: constant temperature's change is less than 0.2 °C /min in the thermocouple testing process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.