Abstract

A surface electromyography (sEMG) signal typically results from the electrical activities of many muscle fibers, and can be utilized as a signal source in prostheses due to its abundance of movement information. This paper proposes an sEMG-detection circuit for the acquisition of the controlling signal in EMG-Bridge (EMGB) systems. The detection circuit mainly comprises a preamplifier, a driven right leg (DRL) circuit, a high-pass filter (HPF), a low-pass filter (LPF), and a gain adjustable amplifying circuit. The common-mode rejection ratio (CMRR) of the circuit is higher than 120 dB, the input impedance is greater than 100 MΩ, the passband range is 20~450 Hz, and the frequency attenuation in stopband is not less than 120dB/dec.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.