Abstract
Minimally invasive surgery has advantage of reducing the size of the body's trauma, post-operative pain and length of hospital stay. This paper proposes a new type of propulsion mechanism that can simulate the specific operations of human hands in vascular interventional procedures. We conducted detailed structural design and analysis of this propulsion mechanism so that the device can perform the function of delivering the catheter and guide wire simultaneously or separately. The new device allows for easy disassembly of the catheter or guide wire and meets the requirements for real-time monitoring of catheter and guide wire delivery resistance during vascular interventional procedures. The advancing mechanism can assist the doctor in completing the basic operations of the vascular intervention surgery and reduce the surgical intensity of the doctor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.