Abstract
In this simulation study, we describe a microstrip transmission line (MTL) array coil, designed to provide a uniform |B1| field along the spinal cord in 7 T MRI. The coil structure was designed by segmenting the spinal cord into four sections according to its shape. The geometry and properties of the MTL coils were optimized individually to obtain a uniform |B1| field distributed according to the shape of the spinal cord in each section. On combining the individually-optimized MTL coils, we present a microstrip array design that produces a highly uniform |B1| field for the whole spinal cord. Furthermore, with this design, the |B1| field contour lines followed the shape of the spinal cord. A comparison with a standard 8-channel loop array coil shows that our design produces a more uniform |B1| field in the target imaging area, demonstrating its potential for spinal cord imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.