For this study, we designed an implantable rectangular spiral antenna for medical biotelemetry in the Medical Implant Communications Service band (402 MHz to 405 MHz). The designed antenna has a U-shaped loop for impedance matching. The antenna impedance is easily adjusted by controlling the shape and length of the U-shaped loop. Significant design parameters were studied to understand their effects on the antenna performance. To verify the potential of the antenna for the desired applications, we fabricated a prototype and measured its performance in terms of the resonant characteristics and gain radiation patterns of the antenna. In the testing phase, the prototype antenna was embedded in human skin tissue-emulating gel, which was developed to simulate a real operation environment. The measured resonant characteristics show good agreement with the simulations, and the -10 dB frequency band is within the range of 398 MHz to 420 MHz. The antenna exhibits a maximum gain of -22.26 dBi and an antenna efficiency of 0.215%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.