Abstract

A remote wireless electrocardiogram (ECG) monitoring system is designed by using the CC2530 micro-controller as the device core. Acquisition, conversion and data processing for ECG signals are realized on CC2530 micro-controllers. And the ECG data is transmitted to the coordinator by using ZigBee. It realizes the real-time monitoring of ECG signals and heart rate variability (HRV) data. The test results show that the maximum error of the designed ECG monitoring analyzer is 3 beats per minute and the average error is 1.6 beats per minute, which can meet the requirement of pharmaceutical industry standards of the People's Republic of China. The ECG monitoring analyzer has good portability, high measurement precision and good practical application values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.