Abstract

Bio-impedance analysis provides non-invasive estimation of body composition. Recently, applications based on bio-impedance measurement in skin tissue such as skin cancer diagnosis and skin composition monitoring have been studied. For scanning the electrical properties along the skin depth, the relationship between the electrode topologies and the depth sensitivity should be clarified. This work reports a systematic analysis on designing line electrode topologies to measure the bio-impedance of the skin layer at specific depth using a finite element method (FEM). Four electrodes consisting of two outer current electrodes and two inner voltage electrodes in the form of Wenner-Schlumberger array were employed on the top of a collagen layer as a skin model. The numerical results demonstrate a change in the effective depth of measurement depending on the electrode topologies, which also have a good agreement with an analytic solution. This study suggests a decision guideline for designing the electrode topologies to achieve target depth sensitivity in bio-impedance measurement of skin tissue.Clinical Relevance-This establishes the effect of electrode topologies on depth sensitivity in bio-impedance measurements in skin layer.