Abstract

We have designed an extended range neutron long counter on the basis of work optimized using SuperMC code. The problem of the existing traditional long counters is that their response function falls rapidly above 5 MeV. We proposed a new designed by adding two layers of converter material inside the polyethylene moderator. The relatively low density chromium and high density lead metals convert high energy neutron by (n, xn) spallation reaction. This produces more neutrons of lower energies, which have higher probability of being detected by thermal 3He-counter. The response function at lower neutron energies was improved by inserting small polyethylene cylinder in front of 3He counter. In this design we achieved to extent the flat response function of the long counter from few keV up to 150 MeV. The total fluctuation of response curve is less than ±9% over the entire energy range. The designed long counter is suitable to be used as neutron monitor for monitoring neutron fluence at high-energy neutron source.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.