Abstract

Single sided Brushless DC (BLDC) drives with axial flux have the advantage of a flat design and high torque. This makes them a good choice for operating less invasive implantable Ventricular Assist Devices (VADs), which can be applied for the therapy of cardio vascular diseases. The disadvantage of such drives is the occurrence of high axial forces. In order to avoid axial forces some VAD drives have air gap windings. Their torque generation completely relies on magnetic stray fluxes. Therefore an increase in efficiency is expected when using iron teeth with single tooth windings. As iron losses have to be considered now, a study is performed to identify the key parameters for iron and copper losses. From this study an efficient VAD drive design is deduced.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.