Abstract

Single-cell analysis of cellular contents by highly sensitive analytical instruments is known as chemical cytometry. A chemical cytometer typically samples one cell at a time, quantifies the cellular contents of interest, and then processes and reports that data. Automation adds the potential to perform this entire sequence of events with minimal intervention, increasing throughput and repeatability. In this chapter, we discuss the design considerations for an automated capillary electrophoresis-based instrument for assay of enzymatic activity within single cells. We describe the key requirements of the microscope base and capillary electrophoresis platforms. We also provide detailed protocols and schematic designs of our cell isolation, lysis, sampling, and detection strategies. Additionally, we describe our signal processing and instrument automation workflows. The described automated system has demonstrated single-cell throughput at rates above 100cells/h and analyte limits of detection as low as 10-20mol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.