Abstract

*In order to design an active-type personal dosemeter having sufficient sensitivity to high-energy neutrons up to 100 MeV, the characteristic response of several types of silicon (Si) semiconductor detectors has been investigated experimentally and theoretically. At first, the response of each detector has been checked to protons accelerated by a Van de Graff accelerator installed at Kobe University. A fully-depleted Si detector was selected as a promising detector element owing to a simple structure and comprehensive characteristics. In the next step, neutron irradiation experiments were carried out in a monoenergetic reference field in JAEA. On the other hand, theoretical distributions of the deposited energy were estimated in a simplified geometry consisting of Si and PE layers. An agreement in the relative shape was confirmed between the theoretical and experimental results. Applying this calculation procedure to other neutron energies and various thickness of sensitive layer, it was found that the energy dependence could be improved by the two-window technique by setting different discrimination levels on the pulse height. Another important characteristic of the angular dependence was also discussed for practical application of a personal dosemeter for high energy neutrons.