Abstract

The moderator structure for a neutron spectrometer was optimized with the Monte Carlo code MCNP-4B. The spectrometer consists of a cylindrical moderator and a position-sensitive thermal neutron detector and obtains energy spectra from thermal neutron distribution along its cylindrical axis. The structure of the moderator was improved by using a low hydrogen density material on one end and a high hydrogen density on the other, and inserting a neutron absorber that eliminates thermal neutron diffusion. This design improves the energy resolution of the spectrometer, especially for low-energy neutrons from a few to 100 keV. The designed spectrometer can be applied to the measurement of energy spectra over a neutron energy range from a few keV to 20 MeV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.