Abstract

A Rogowski coil is developed to detect the nanosecond pulse signals of the discharge current with a wide bandwidth of 800 kHz to 106 MHz and high sensitivity of 2.22 V/A. Performance tests show that the Rogowski coil has both excellent dynamic and static characteristics. Calibrating results and the comparison between the standard current shunt and the developed Rogowski coil for the measurement of nanosecond discharge pulses demonstrate that the developed Rogowski coil can reproduce the actual waveform of the discharge current accurately.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.