Abstract

Catheterization remains the "gold standard" for bladder volume assessment, but it is invasive and introduces the risk of infections and traumas. Therefore, noninvasive bladder volume measurement methods have gained interest. In a preceding study a new technique to measure the bladder volume on the basis of nonlinear ultrasound wave propagation was validated. This paper describes a first prototype of a dedicated multilayer transducer to implement this approach. It is composed of a PZT transducer for transmission and a PVDF layer for reception. Acoustical measurements in a water tank and phantom measurements showed that there is a relation between bladder volume and the harmonic contents of the echo obtained from a region of interest behind the bladder. Simulations with an equivalent transducer model on the basis of KLM-circuit modeling closely matched with the results from the acoustical measurements. The results demonstrated the feasibility of the multilayer transducer design for bladder volume assessment on the basis of nonlinear wave propagation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.