Abstract

Optical coherence tomography (OCT) has a wide range of uses in bioimaging and nondestructive testing. Larger bandwidth light sources have recently been implemented to enhance measurement resolution. Increased bandwidth has a negative impact on spectral nonlinearity in k space, notably in the case of spectral domain OCT (SD-OCT). This nonlinearity reduces the depth-dependent signal sensitivity of the spectrometers. A grating and prism combination is extensively used for linearizing. In an earlier study, we used a combination of the reflective grating and prism, as well as a cylindrical mirror with a radius of 180 mm, to achieve a high SR ratio with low nonlinearity. A creative design for a spectrometer with a cylindrical mirror of radius 50 mm, a light source with a center wavelength of 830 ± 100 nm (μm−1 − 6.756 μm−1 in k-space), and a grating of 1600 lines/mm is presented in this work. The design optimization is performed using MATLAB and ZEMAX. In the proposed design, the nonlinearity error reduced from 157 μm to 10.75 μm within the wavenumber range considered. The sensitivity research revealed that, with the new design, the SR ratio is extremely sensitive to the imaging optics’ angles. To resolve this, a spectrometer based on Grism is introduced. We present a Grism-based spectrometer with an optimized SR ratio of 0.97 and nonlinearity of 0.792 μm (). According to the sensitivity study, the Grism-based spectrometer is more robust.