Abstract
The sensitivity of linear variable differential transformer (LVDT) position sensors to external slowly varying magnetic fields represents a critical issue when these sensors are installed close to high-current cables or electrical motors with significant fringe fields. The resulting position error can reach several hundreds of micrometers against a specified uncertainty normally below a few micrometers. In this paper, the design of a LVDT position sensor with high rejection to external constant or slowly varying magnetic fields is addressed by exploiting the finite element method (FEM) simulator FLUX. A shield, isolated from the sensor's magnetic circuit, has been considered to reduce the effect of magnetic fields on the secondary voltages of the LVDT. In addition, a dc current is used in order to polarize the magnetic circuit to reduce the sensitivity of the sensor to external interferences.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.