Abstract

ABSTRACTWe discuss the design of a new backscattering spectrometer to be installed at the Cold Neutron Research Facility at the National Institute of Standards and Technology. Si (111) crystals cover both monochromator and analyzer which are spherically bent to a radius of curvature of ~ 2 m to focus the incident and scatterered neutron beams. The bending increases the intrinsic lattice gradient of Si beyond its Darwin limit, resulting in an energy resolution of ~ 0.75 μeV FWHM. The monochromator is Doppler-driven, allowing users access to a dynamic range of ±60 μeV. The elastic Q-range covers 0.15 to 1.8 Å-1. The most novel aspect of this design lies in the incorporation of a phase-space-transform chopper. This device rotates at 4700 rpm while neutrons are Bragg-diffracted from sets of pyrolytic graphite crystals affixed to its periphery. The process enhances the neutron flux at the backscattering energy of 2.08 meV, but at the expense of a larger horizontal divergence. Computer simulations indicate a resultant flux increase of order 3 should be obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.