Abstract

Studies of osteoarthritis initiation and progression that measure strain in cartilage require physiological loading levels. Many studies use magnetic resonance (MR) imaging, which necessitates a MR-compatible loading device. In this study, the design and validation of a new device, the cartilage compressive actuator (CCA), is presented. The CCA is designed for high-field (e.g., 9.4 T) small-bore MR scanners, and meets a number of design criteria. These criteria include capability for testing bone-cartilage samples, MR compatibility, constant load and incremental strain application, a water-tight specimen chamber, remote control, and real time displacement feedback. The mechanical components in the final design include an actuating piston, a connecting chamber, and a sealed specimen chamber. An electro-pneumatic system applies compression, and an optical Fibre-Bragg grating (FBG) sensor provides live displacement feedback. A logarithmic relationship was observed between force exerted by the CCA and pressure (R2 = 0.99), with a peak output force of 653 ± 2 N. The relationship between FBG sensor wavelength and displacement was linear when calibrated both outside (R2 = 0.99) and inside (R2 = 0.98) the MR scanner. Average slope was similar between the two validation tests, with a slope of −4.2 nm/mm observed inside the MR scanner and −4.3 to −4.5 nm/mm observed outside the MR scanner. This device meets all design criteria and represents an improvement over published designs. Future work should incorporate a closed feedback loop to allow for cyclical loading of specimens.