Abstract
Rotating Modulation Collimators (RMC) are devices that can be used to determine the location of a radiation source. In this work, a design with extended field-of-view (FOV) is proposed to allow faster measurements by providing a wider scanned area. Monte Carlo simulations were used to calculate the response of the RMC system, system matrix and modulation profiles. The Maximum Likelihood Expectation Maximization algorithm was used for image reconstruction. The proposed design provides a FOV up to 60° with a maximum error of 5° or up to 70° with a maximum error of 15°.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.