Abstract
Rehabilitation medicine studies have indicated that occupational therapy has a positive effect on the recovery of hand function. However, few wearable hand exoskeletons can assist in performing the adduction and abduction exercises of occupational therapy. Therefore, the wearable index finger rehabilitation exoskeleton (WIFRE) is studied in this paper. First, a WIFRE is proposed to realize the index joint’s independent actuation, including adduction and abduction movements. Second, the local/global kinematic and dynamic manipulability measures are proposed and analyzed to evaluate the performance of the WIFRE. The analysis results show that the dimensional parameters of the WIFRE have a significant effect on its global manipulability measures. The global kinematic and dynamic manipulability measures can be improved by 13%–15% compared with the corresponding minimums. Third, a multi-parameter multi-objective optimization method is proposed to simultaneously enhance the three global manipulability measures. Finally, experiments are performed to verify the effectiveness of the index finger exoskeleton and the global manipulability measures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.