Abstract

A high resolution x-ray imaging system designated the solid state x-ray image intensifier (SSXII) is being developed as a potential replacement of conventional lower-resolution x-ray detectors for neurovascular interventions. The SSXII consists of a CsI(Tl) phosphor to convert x-rays to light, a fiber optic taper to extend the field-of-view and to couple the phosphor to an Electron-Multiplying CCD (EMCCD) which has a built-in gain to boost the signal above the read-out noise floor. A parallel cascade linear-system model is explored and applied to quantitatively analyze the signal and noise transfer at each stage, and optimize the detector performance. The modulation transfer function (MTF), detective quantum efficiency (DQE) and frequency dependent instrumentation noise equivalent exposure (INEE) can be obtained from the theoretical model and experimental measurement. Agreement of these results validates the effectiveness of the model. Finally, compared with the traditional flat-panel detector (FPD), the SSXII exhibits better performance with higher spatial resolution, higher DQE and lower INEE.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.