Abstract
This study investigates the feasibility of employing an in-house-developed physical phantom dedicated to the weekly quality control testing of radiographic systems, performed by radiographers. For this purpose, a 3D phantom was fabricated, featuring test objects, including a model representing a lesion. Alongside this phantom, a commercial phantom, specifically, IBA's Primus L, was utilized. Weekly imaging of both phantoms was conducted over a span of four weeks, involving different imaging protocols and anode voltages. Subsequently, the obtained data underwent visual evaluation, as well as measurement of the intensity of selected regions of interest. The average values for three incident kilovoltages remained consistently stable over the four weeks, with the exception of the "low energy" case, which exhibited variability during the first week of measurements. Following experiments in "Week 1", the X-Ray unit was identified as malfunctioning and underwent necessary repairs. The in-house-developed phantom demonstrated its utility in assessing the performance of the X-Ray system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.