Abstract

We present the design, construction and testing of a microelectrode array with spatial resolution which can be used for the detection of cancerous cells in a mixed cell culture by means of the impedance spectroscopy technique. Two different microelectrode diameters, namely 57 m and 211 m, were tested. A layer of an AuCu alloy was deposited between the glass substrate and the exterior Au layer, enhancing the adhesion of the microelectrodes to the substrate and at the same time showing lower impedance than the commonly used material Au. Characteristic features were extracted from the impedance measurements at different phases (growth, confluence, wound and healing) and used to feed a Linear Discriminant Analysis algorithm in order to discriminate between normal and cancerous cells. Relevant statistical tests were applied in the discrimination model for each phase. Finally, it was determined that the larger microelectrodes have a superior discriminant capacity: no incorrect classifications were obtained with microelectrodes measuring 211 m in diameter, while 23.5% false positives and 5.6% false negatives were obtained with 57 m microelectrodes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.