Cyclic flexure and stretch are essential to the function of semilunar heart valves and have demonstrated utility in mechanically conditioning tissue-engineered heart valves. In this study, a cyclic stretch and flexure bioreactor was designed and tested in the context of the bioresorbable elastomer poly(glycerol sebacate). Solid poly(glycerol sebacate) membranes were subjected to cyclic stretch, and micromolded poly(glycerol sebacate) scaffolds seeded with porcine aortic valvular interstitial cells were subjected to cyclic stretch and flexure. The results demonstrated significant effects of cyclic stretch on poly(glycerol sebacate) mechanical properties, including significant decreases in effective stiffness versus controls. In valvular interstitial cell-seeded scaffolds, cyclic stretch elicited significant increases in DNA and collagen content that paralleled maintenance of effective stiffness. This work provides a basis for investigating the roles of mechanical loading in the formation of tissue-engineered heart valves based on elastomeric scaffolds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.