Abstract

The Coronavirus Disease 2019 (COVID-19) pandemic renewed interest in infectious aerosols and reducing risk of airborne respiratory pathogen transmission, prompting development of devices to protect healthcare workers during airway procedures. However, there are no standard methods for assessing the efficacy of particle containment with these protective devices. We designed and built an aerosol bio-containment device (ABCD) to contain and remove aerosol via an external suction system and tested the aerosol containment of the device in an environmental chamber using a novel, quantitative assessment method. The ABCD exhibited a strong ability to control aerosol exposure in experimental and computational fluid dynamic (CFD) simulated scenarios with appropriate suction use and maintenance of device seals. Using a log-risk-reduction framework, we assessed device containment efficacy and showed that, when combined with other protective equipment, the ABCD can significantly reduce airborne clinical exposure. We propose this type of quantitative analysis serves as a basis for rating efficacy of aerosol protective enclosures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.